Microprocessor correct the value of electromotive force of saturated standard cell
Keywords:
saturated normal element, error, standard current controlled stabilizer, thermo–adjustor.Abstract
The temperature corrections of errors is analysed and their dependence on temperature measurement errors of the normal element is defined. It is shown that in the case of measurement the temperature deviation of the normal element from the calibration temperature higher accuracy of the amendments values determination is achieved. The structure of the thermo–emf microprocessor measure based on saturated standard cell type H482 with automatic temperature correction of the thermo–emf values is proposed. The method of processing the results of temperature deviation, which fully compensates the effect of offset voltage of operational amplifiers and analog–digital converter. We describe a digital linearization method of transfer characteristics of the thermoresistive converter. The thermo–adjustor structure is designed and the main results of experimental studies of the thermo–emf microprocessor measure are presented.