THE PAPER MASS SCAPULAR CONCENTRATION METER ACCURACY IMPROVE ALGORITHMICAL METHOD

Authors

  • О. М. Романюк Національний університет «Львівська політехніка», 79013, м. Львів, вул. С. Бандери, 12
  • Б. А. Кріль Національний університет «Львівська політехніка», 79013, м. Львів, вул. С. Бандери, 12
  • О. В. Кріль Національний університет «Львівська політехніка», 79013, м. Львів, вул. С. Бандери, 12

Keywords:

blade consistometer, blade consistency meter, paper pulp, consistency, non-Newtonian fluid.

Abstract

The algorithmic method of precision enhancement of paper pulp blade consistometers is considered. In the said method the current strength in the winding of a moving coil of a magnetoelectric transducer in the steady regime of sensor`s blade motion condition is measured during its movement in downstream and upstream directions. By processing this data the errors, caused by the influence of the moving medium on the consistometer sensitive blade, can be considered and excluded. Schematic diagram of a paper pulp consistometer with paper pulp flow velocity compensation is presented. Dependence of voltage and current versus time on the outputs of the main blocks of presented paper pulp consistometer with compensation of paper pulp flow velocity influence is considered. Iinfluence of paper pulp flow velocity and consistency on the dependencies of current versus time for the coil of a magnetoelectric transducer of a paper pulp consistometer is described.

Downloads

Download data is not yet available.

References

1. Waller, M.H. (1993) A tutorial on the measurement of paper stock consistency. ISA Transactions, 32, pp.277-282.
2. Технология целлюлозно – бумажного производства; у 3 т. Т. 1. Сирье и производство полуфабрикатов/ С.С Пузырев, Є.В. Виролайнен, Ю.А. Поляков, А.М. Кряжев. – СПб.: Политехника, 2004. – 316 с.
3. Schramm, G. A. (2000) Practical Approach to Rheology and Rheometry. 2nd Edition, Gebrueder HAAKE GmbH, Karlsruhe, Federal Republic of Germany, 291 p.
4. Steffe, J.F. (1996) Rheological Methods in Food Process Engineering. 2nd Edition. Freeman Press, East Lansing, MI, USA 418p.
5. Valmet Blade Consistency Transmitter – Valmet SP Installation and User's Manual, Valmet Automation Inc., Kajaani, Finland. – 2015.
6. Metso Blade Consistency Transmitter. Installation and User's Manual K11389 V1.1 EN Metso Automation Inc. Documentation: Metso Automation / Marjo Nygård Copyright © December 2012.
7. Wall, H. A. (1968) Consistency measuring apparatus. US Patent No US 3364730 A.
8.“MBT-2500 User manual” (2007) BTG Pulp and Paper Technology AB, Säffle, Sweden.
9. Skallen, B. (1989) Method and apparatus for measuring the viscosity of a medium. US Patent No US4875362.
10. Романюк О.М., Кріль Б.А. Підвищення точності ротаційних віскозиметрів алгоритмічним шляхом //Матеріали ІІ Міжнародної науково-технічної конференції «Сучасні методи, інформаційне, програмне та технічне забезпечення систем управління організаційно-технічними та технологічними комплексами». – К: НУХТ, 2015 р.

Published

2017-03-25

How to Cite

Романюк, О. М., Кріль, Б. А., & Кріль, О. В. (2017). THE PAPER MASS SCAPULAR CONCENTRATION METER ACCURACY IMPROVE ALGORITHMICAL METHOD. METHODS AND DEVICES OF QUALITY CONTROL, (1(38), 28–33. Retrieved from https://mpky.nung.edu.ua/index.php/mpky/article/view/359

Issue

Section

METHODS AND EQUIPMENT OF NON-DESTRUCTIVE CONTROL